1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 120 121 122 123 124 125 126 127 128 129 130 131 132 133 134 135 136 137 138 139 140 141 142 143 144 145 146 147 148 149 150 151 152 153 154 155 156 157 158 159 160 161 162 163 164 165 166 167 168 169 170 171 172 173 174 175 176 177 178 179 180 181 182 183 184 185 186 187 188 189 190 191 192 193 194 195 196 197 198 199 200 201 202 203 204 205 206 207 208 209 210 211 212 213 214 215 216 217 218 219 220 221 222 223 224 225 226 227 228 229 230 231 232 233 234 235 236 237 238 239 240 241 242 243 244 245 246 247 248 249 250 251 252 253 254 255 256 257 258 259 260 261 262 263 264 265 266 267 268 269 270 271 272 273 274 275 276 277 278 279 280 281 282 283 284 285 286 287 288 289 290 291 292 293 294 295 296 297 298 299 300 301 302 303 304 305 306 307 308 309 310 311 312 313 314 315 316 317 318 319 320 321 322 323 324 325 326 327 328 329 330 331 332 333 334 335 336 337 | <!DOCTYPE html PUBLIC "-//W3C//DTD HTML 4.01 Transitional//EN">
<html>
<head>
<title>illumread</title>
<meta http-equiv="content-type" content="text/html;
charset=windows-1252">
<meta name="author" content="Graeme Gill">
</head>
<body>
<h2><b>spectro/illumread</b></h2>
<h3>Summary</h3>
Use an instrument or instruments to measure an illuminant spectrum,
including estimate its Ultra Violet content. A combination of direct
illumination readings and readings from a piece of paper having some
FWA content are used for this. (If the UV content is not needed, or
a suitable instrument is not available, then <a
href="spotread.html">spotread</a> should be used instead.)<br>
<h3>Usage Summary</h3>
<small><span style="font-family: monospace;">illumread [-options]
illuminant.sp</span><br style="font-family: monospace;">
<span style="font-family: monospace;"> </span><a
style="font-family: monospace;" href="#v">-v</a><span
style="font-family: monospace;">
Verbose mode</span><span style="font-family:
monospace;"></span></small><small><span style="font-family:
monospace;"></span></small><br style="font-family: monospace;">
<small><span style="font-family: monospace;"></span><span
style="font-family: monospace;"> </span><a
style="font-family: monospace;" href="#S">-S</a><span
style="font-family: monospace;">
Plot the readings in a graph window.</span><br
style="font-family: monospace;">
<span style="font-family: monospace;"> </span><a
style="font-family: monospace;" href="#c">-c comport</a><span
style="font-family: monospace;">
Set
COM port, 1..4 (default 1)</span><span style="font-family:
monospace;"></span><span style="font-family: monospace;"></span><span
style="font-family: monospace;"></span></small><small><span
style="font-family: monospace;"></span></small><font size="-1"><span
style="font-family: monospace;"></span></font><font size="-1"><span
style="font-family: monospace;"><br>
</span></font><font size="-1"><span style="font-family:
monospace;"> <a href="#N">-N</a>
Disable
initial calibration of instrument</span></font> if possible<br>
<font size="-1"><span style="font-family: monospace;"> </span><a
style="font-family: monospace;" href="#H">-H</a><span
style="font-family: monospace;">
Use high resolution spectrum mode
(if available)<br>
<font size="-1"> <a href="#Yr">-Y r</a>
Set refresh measurement mode</font><br>
</span></font><font size="-1"><span style="font-family:
monospace;"> <a href="#W">-W n|h|x</a>
Override
serial port flow control: n = none, h = HW, x = Xon/Xoff</span></font><br>
<small><span style="font-family: monospace;"> </span><a
style="font-family: monospace;" href="#T">-T</a><span
style="font-family: monospace;">
Test mode - restore & save measurements to<br>
*_i.sp, *_r.sp, *_p.sp, *_mpir.sp, *_cpir.sp files<br>
</span></small> <small><span style="font-family: monospace;"> </span><a
style="font-family: monospace;" href="#D">-D [level]</a><span
style="font-family: monospace;">
Print debug diagnostics to stderr</span></small><br>
<font size="-1"><span style="font-family: monospace;"><a
href="#file"><span style="font-style: italic;">illuminant.sp</span></a>
File
to save measurement to<br style="font-family: monospace;">
</span></font><small><span style="font-family: monospace;"></span><span
style="font-family: monospace;"></span></small><br>
<h3>Usage Details and Discussion</h3>
<b>illumread</b> uses a suitable instrument to read an illuminant
spectrum, and uses an indirect method to estimate the Ultra Violet
content of the illuminant, so as to provide better accuracy with <a
href="FWA.html">FWA compensation</a>. An instrument or combination
of instruments capable of spectral measurement of both emissive
measurement and reflective measurement without a U.V. filter is
required for this.<br>
<br>
<a name="v"></a>The <b>-v</b> flag causes extra information to be
printed out during chartread operation.<br>
<br>
<a name="S"></a>The <b>-S</b> flag enables the plotting of the
spectral reflectance/transmittance values. You must select the plot
window and strike a key in it to continue with another measurement.<br>
<br>
<a name="c"></a> The instrument is assumed to communicate through a
USB or serial communication port, and the initial port can be
selected with the <b>-c</b> option, if the instrument is not
connected to the first port. If you invoke <span
style="font-weight: bold;">illumread</span> so as to display the
usage information (i.e. "illumread -?" or "illumread --"), then the
discovered USB and serial ports will be listed. On UNIX/Linux, a
list of all possible serial ports are shown, but not all of them may
actually be present on your system.<br>
<br>
<a name="N"></a><span style="font-weight: bold;">-N</span> Any
instrument that requires regular calibration will ask for
calibration on initial start-up. Sometimes this can be awkward if
the instrument is being mounted in some sort of measuring jig, or
annoying if several sets of readings are being taken in quick
succession. The -<span style="font-weight: bold;">N</span>
suppresses this initial calibration if a valid and not timed out
previous calibration is recorded in the instrument or on the host
computer. It is advisable to only use this option on the second and
subsequent measurements in a single session.<br>
<br>
<a name="H"></a> The -<span style="font-weight: bold;">H</span>
option turns on high resolution spectral mode, if the instrument
supports it. See <a href="instruments.html">Operation of particular
instruments</a> for more details.<br>
<br>
<a name="Yr"></a> The -<span style="font-weight: bold;">Y r</span>
option turns on refresh mode measurement, if the instrument supports
it. This may improve the repeatability of measurements of
illuminants that have a repetitive flicker.<br>
<br>
<a name="W"></a>The <b>-W</b> <span style="font-weight: bold;">n|h|x</span>
parameter overrides the default serial communications flow control
setting. The value <span style="font-weight: bold;">n</span> turns
all flow control off, <span style="font-weight: bold;">h</span>
sets hardware handshaking, and <span style="font-weight: bold;">x</span>
sets Xon/Xoff handshaking. This commend may be useful in workaround
serial communications issues with some systems and cables. <br>
<br>
<a name="T"></a>The <b>-T</b> flag invokes the test mode. In test
mode the three measurements are saved to files <i>illuminant</i><i>_i.sp</i>
(Illuminant spectrum), <i>illuminant</i><i>_r.sp</i> (Illuminant
off paper spectrum), and <i>illuminant_p.sp (</i>Instrument
measured paper reflectance spectrum), and these will be loaded if
discovered, allowing a replay of the calculation without requiring
any measurement. In addition, two diagnostic files <i>illuminant</i><i>_mpir.sp</i>
(Measured paper under illuminant spectrum) and <i>illuminant</i><i>_cpir.sp
(</i>Computed paper under illuminant spectrum) will be saved.<br>
<br>
<a name="D"></a>The <b>-D</b> flag causes communications and other
instrument diagnostics to be printed to stdout. A level can be set
between 1 .. 9, that may give progressively more verbose
information, depending on the instrument. This can be useful in
tracking down why an instrument can't connect.<br>
<br>
<a name="file"></a>The <span style="font-weight: bold; font-style:
italic;">illuminant.sp</span> is the name of the file to save the
resulting illuminant spectrum to. The format used is <a
href="File_Formats.html#.sp">.sp</a>.<br>
<br>
<hr style="width: 100%; height: 2px;"><br>
Unlike the other measurement utilities, <span style="font-weight:
bold;">illumread</span> doesn't connect to the instrument until it
is about to make a measurement. This allows for the possibility of
using a different instrument for each measurement.<br>
<br>
It will display a menu:<br>
<br>
Press 1 .. 6<br>
1) Measure direct illuminant<br>
2) Measure illuminant reflected from paper<br>
3) Measure paper<br>
4) Select another instrument, Currently 1 'usb:/bus4/dev2/
(GretagMacbeth i1 Pro)'<br>
5) Compute illuminant spectrum, average result with 0 previous
readings & save it<br>
6) Compute illuminant spectrum from this reading & save result<br>
7) Exit<br>
<br>
There are three measurements to be made, after which the illuminant
can be computed and saved. Before each measurement, the instrument
may need calibrating.<br>
<br>
The first measurement needs a spectral instrument capable of reading
in an ambient or emissive mode. For instance, a Spectrolino, Eye-One
Pro or ColorMunki would be suitable instruments.<br>
<br>
The second measurement needs a spectral instrument capable of
reading in an projector or emissive mode. For instance, a
Spectrolino, Eye-One Pro or ColorMunki would be suitable
instruments.<br>
<br>
The third measurement needs a spectral instrument capable of reading
in reflective mode with UV included. For instance, a Spectrolino,
Eye-One Pro, DTP20, DTP22 or DTP41 would be suitable
instruments, as long as they are not fitted with UV filters.<br>
<br>
To be able to estimate the level of Ultra Violet (UV) light in the
illuminant, a reasonable sized piece of white paper needs to be
used. The paper should have some noticeable level of FWA
(Fluorescent Whitener Additive, or Optical Brightening Agents) in
it, so that it responds to UV light. A piece of cheap copier paper
is ideal, since cheap paper is typically whitened with large amounts
of FWA. If the paper is thin (less than 160 gsm) then two or three
sheets should be used to prevent any background showing through. If
the intention is to use the illuminant spectrum for proofing to a
particular paper, then it's best to use the intended paper for this
purpose.<br>
<br>
The first measurement <span style="font-weight: bold;">1)</span>,
is to use either the ambient or emissive measurement mode to measure
the illumination directly.<br>
<br>
<div style="margin-left: 40px;">If the instrument supports an
ambient measurement capability, then it will be used. If the
insrument does not have an ambient mode, then an emissive
measurement mode can be used, although typically many illuminants
are too bright to directly point the instrument at. A work-around
is to reflect the illuminant from a spectrally flat white surface.
A good candidate for this is a piece of white, fine textured
polystyrene foam. [The suitability of a reflector can be checked
using <span style="font-weight: bold;">spotread -S</span> to
check that the reflection characteristic is close to flat.]<br>
<br>
<img alt="Measuring Ambient" src="illumread_1.jpg" width="143"
height="272"> <img style="width: 141px; height: 282px;"
alt="Measuring Ambient" src="illumread_2.jpg"> <img
style="width: 226px; height: 282px;" alt="Measuring Ambient"
src="illumread_3.jpg"><br>
</div>
<br>
The second measurement <span style="font-weight: bold;">2)</span>,
is to measure the illuminant after it has reflected from the paper.<br>
<br>
<div style="margin-left: 40px;">This is done by placing the paper
such that it is uniformly illuminated with reasonable brightness,
and then placing the instrument so that it receives the reflected
light from the paper. This is typically achieved by placing the
instrument close to the paper at about 45ยบ, so that it's aperture
has a clear view of the illuminated paper, but avoiding shadowing
the region that is in view, and also avoiding measuring specular
reflection if the paper is glossy.<br>
<br>
<img style="width: 219px; height: 261px;" alt="Measuring via
Paper" src="illumread_5.jpg"><img style="width: 252px; height:
259px;" alt="Measuring via Paper" src="illumread_4.jpg"><br>
</div>
<br>
The third measurement <span style="font-weight: bold;">3)</span>,
is to measure the paper directly using the instrument reflective
mode measurement.<br>
<div style="margin-left: 40px;"><img style="width: 186px; height:
162px;" alt="Measuring Paper" src="illumread_6.jpg"><br>
</div>
If a different instrument is needed, use <span style="font-weight:
bold;">4)</span> to select from the available instruments attached
to your computer.<br>
<br>
Once these three measurements have been made, then the illuminant
readings spectrum can be computed and save using <span
style="font-weight: bold;">6)</span>, or a series of readings can
be made with each reading being averages with the previous readings
before saving it by using <span style="font-weight: bold;">5)</span>.
Note that the averaged readings will be weighted by their absolute
intensities, and that while the direct and indirect illumination
needs measuring for each reading, the same paper measurement can be
used each time.<br>
<br>
If plotting is enabled, a plot of the measured (black) and with
estimated UV (red) is plotted. This is followed by a plot showing
measured paper reflectance (black) and the FWA calculated paper
reflectance (red).<br>
<br>
<br>
Illumread can then be terminated using <span style="font-weight:
bold;">7)</span>.<br>
<br>
<br>
<br>
<br>
<br>
<br>
<br>
<br>
<br>
<br>
<br>